CryoPro Cryo Surgical Flask Frequently asked Questions

Occasionally you may be concerned about how your cryo Flask is sounding or looking please see the FAQ's to help alleviate andy issues.

CryoPro Cryo Surgical Flask Frequently asked Questions

Why is the spray function restricted or the valve freezing?

Observation

1) No or restricted spray function. No or restricted spray will be noticed when removing the spray aperture and pulling the trigger.
OBS! Lift the top of the pressure relief valve to distinguish between a clogged main valve and loss of pressure (see below).

2) The valve freezes up and sticks open (leaks).

Explanation

Clogging of the main valve is a result of contamination of the storage tank or the bottle itself.
1) Would normally be caused by and accumulation of impurities or ice crystals.
2) strongly indicates water/ice in the valve.

Fix

Return the unit for thorough cleaning of the main valve.

Prevention

- Clean the CryoPro can: fill it with warm water and a few drops of dish soap, clean with a soft brush, empty and dry the bottle thoroughly.
- Do not leave the CryoPro open when empty. Leave the top on.
- Clean the storage tank regularly: empty it completely before refilling. Use a vacuum cleaner for 20 – 30 min. to create air circulation in the empty tank to dry out water from melting ice crystals (the hose should be positioned approx. half ways into the tank).

Why is the spray function restricted?

Why is there a "cold spot" on the bottle?

Why does the top cover condensate and drip?

How to open/refill a pressurized unit?

How to fix a hard and noisy top cover?

How to fix a lack of spray function?

Why does the pressure relief valve make a hissing sound?

Observation

Why is the spray function intermittent?

What to do when the bottle frosts over?